RKI Instruments, Inc. • 1855 Whipple Rd. • Hayward, CA 94544 • (510) 441-5656
Instruction Manual
Eagle Series
Portable Multi-Gas Detector
Part Number: 71-0028RK
Edition: Third
Released: July 2001
Zobrazit stránku 0
1 2 3 4 5 6 ... 79 80

Shrnutí obsahu

Strany 1 - Eagle Series

RKI Instruments, Inc. • 1855 Whipple Rd. • Hayward, CA 94544 • (510) 441-5656Instruction ManualEagle SeriesPortable Multi-Gas DetectorPart Number: 71-

Strany 2 - Warranty

4 • Specifications Eagle Instruction ManualTable 2 lists specifications for the Eagle’s standard sensors. Your Eagle model may not include all of the s

Strany 3 - Table of Contents

Eagle Instruction Manual Description • 5DescriptionCaseThe Eagle has a plastic case with a full-sized handle. The high-visibility case is shielded to

Strany 4

6 • Description Eagle Instruction ManualAlarm LightsTwo ultra-bright, red, light-emitting diodes (LEDs) provide visual alarms for gas concentrations a

Strany 5

Eagle Instruction Manual Description • 7SensorsThis section describes the Eagle’s standard sensors. Non-standard sensors are described in Appendices C

Strany 6

8 • Description Eagle Instruction ManualCircuit BoardsThe Eagle circuit boards analyze, record, control, store, and display the information collected.

Strany 7 - Introduction

Eagle Instruction Manual Operation • 9OperationThe Eagle has four operating modes: normal operating mode, display mode, setup mode, and calibration mo

Strany 8 - About this Manual

10 • Operation Eagle Instruction ManualThe Battery Voltage screen displays the minimum usable and actual battery voltage (for example, 6.0V). If the b

Strany 9

Eagle Instruction Manual Operation • 11CAUTION:Do not use gas from a cigarette lighter to test response to combustibles. Exposing the combustible gas

Strany 10

12 • Operation Eagle Instruction ManualMonitoring Combustible Gases Other than MethaneIf the combustible gas sensor is calibrated to methane (CH4), us

Strany 11 - Description

Eagle Instruction Manual Operation • 13Setting User AccessThe CAL/SETUP switch controls the Eagle functions available to the user. The switch setting

Strany 12 - Sample-Drawing System

Warranty Eagle Instruction ManualWarrantyRKI Instruments, Inc., warranties gas alarm equipment manufactured by RKI and sold by RKI to be free from def

Strany 13 - Oxygen sensor

14 • Alarms Eagle Instruction ManualAlarmsAlarm IndicationsThis section describes the Eagle’s audible and visual alarm indications for gas, over range

Strany 14 - CAL/SETUP Switch

Eagle Instruction Manual Alarms • 15STEL alarm (toxics only)If a toxic gas channel’s average gas reading for the past 15 minutes exceeds the STEL alar

Strany 15 - Operation

16 • Alarms Eagle Instruction ManualLow flow alarmIf the Eagle’s sample system becomes restricted or blocked (for example plugged probe, fouled filter,

Strany 16

Eagle Instruction Manual Alarms • 17Sensor failure alarm and emergency operationThe Eagle continuously monitors itself for proper operation. If a malf

Strany 17 - Normal Operation

18 • Alarms Eagle Instruction ManualResetting Gas AlarmsYou can set the Eagle’s gas alarms for latching or self-resetting alarms(see “Updating the Ala

Strany 18

Eagle Instruction Manual Display Mode • 19Display ModeThe Eagle has four operating modes: normal operating mode, display mode, setup mode, and calibra

Strany 19 - Turning Off the Eagle

20 • Display Mode Eagle Instruction ManualTo enter a user and station ID:To scroll to the next screen at any time, press the DISP/ADJ button.1. Press

Strany 20 - Alarm Indications

Eagle Instruction Manual Display Mode • 21TWA/STEL ScreenThe TWA/STEL screen displays the time-weighted average (TWA) and the short-term exposure limi

Strany 21 - Over range alarm

22 • Display Mode Eagle Instruction ManualClear Data Logger ScreensCAUTION:Once you clear the data logger, you cannot retrieve any data previously sto

Strany 22 - Low battery alarm

Eagle Instruction Manual Setup Mode • 23Setup ModeNOTE: The screens illustrated in this section are examples only. The screens displayed by your Eagle

Strany 23

Eagle Instruction Manual Table of ContentsTable of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 24 - Resetting Gas Alarms

24 • Setup Mode Eagle Instruction Manual• To exit setup mode, from the main menu place the prompt next to the last menu option, START MEASUREMENT, the

Strany 25 - Display Mode

Eagle Instruction Manual Setup Mode • 25Updating Channel SettingsThis procedure describes how to update channel settings for the combustible gas, oxyg

Strany 26 - Elapsed Time Screen

26 • Setup Mode Eagle Instruction ManualThe number in parenthesis indicates the display increment for that portion of the PPM range. In the example be

Strany 27 - Date/Time Screen

Eagle Instruction Manual Setup Mode • 27The maximum fullscale setting for the combustible gas channel is 50,000 PPM; the minimum setting is 1000 ppm.

Strany 28 - Remaining Log Time Screen

28 • Setup Mode Eagle Instruction ManualUpdating oxygen channel settingsThis section describes how to update the target gas label, fullscale setting,

Strany 29 - Setup Mode

Eagle Instruction Manual Setup Mode • 29Updating the display increment setting1. Press the SHIFT/▼ button to place the prompt in the third line, then

Strany 30 - Entering Setup Mode

30 • Setup Mode Eagle Instruction Manual5. Press the AIR/▲ or SHIFT/▼ buttons to display the available target gas labels for the toxic gas channel (H2

Strany 31 - Updating Channel Settings

Eagle Instruction Manual Setup Mode • 31Updating the display increment setting1. Press the SHIFT/▼ button to place the prompt in the third line, then

Strany 32

32 • Setup Mode Eagle Instruction ManualUpdating the Alarm Point SettingsEach of the Eagle’s gas detection channels includes low and high gas alarms.

Strany 33 - Equivalent to 100% LEL

Eagle Instruction Manual Setup Mode • 337. To exit the Set Alarm Points menu, press the SHIFT/▼ button until the prompt is next to Channel 4, then pre

Strany 34 - Updating the target gas label

Table of Contents Eagle Instruction ManualDisplay Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19User an

Strany 35

34 • Setup Mode Eagle Instruction Manual2. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.3. Press the POWER/ENTER button to enter t

Strany 36

Eagle Instruction Manual Setup Mode • 35Turning the User ID Function On or OffWith User ID Input ON, the User and Station ID screen displays during st

Strany 37

36 • Setup Mode Eagle Instruction Manual2. Press and hold the SHIFT/▼ button, then press the DISP/ADJ button. The Auto Calibration screen for the comb

Strany 38

Eagle Instruction Manual Setup Mode • 37With Auto Fresh Air Adjust ON, the Eagle automatically set the fresh air reading for all channels during the s

Strany 39

38 • Setup Mode Eagle Instruction Manual1. From the main menu, select the LOG DATA OVER WRITE menu option.2. Press the AIR/▲ or SHIFT/▼ button to disp

Strany 40

Eagle Instruction Manual Setup Mode • 393. Press the POWER/ENTER button to enter the setting. The day setting flashes.4. Repeat steps 2 and 3 to enter

Strany 41

40 • Setup Mode Eagle Instruction ManualTo reset all default settings:1. From the main menu, select the DEFAULT menu option.2. Press the POWER/ENTER b

Strany 42

Eagle Instruction Manual Calibration • 41CalibrationCalibrate the Eagle when you replace a sensor. Also calibrate the Eagle periodically to assure pro

Strany 43

42 • Calibration Eagle Instruction ManualCalibrating the EaglePress and hold the SHIFT/▼ button, then press the DISP/ADJ button. The Calibration menu

Strany 44

Eagle Instruction Manual Calibration • 433. To adjust the values on the screen, hold down the SHIFT/▼ button, and press the DISP/ADJ button. The Auto

Strany 45 - Returning to Default Settings

Eagle Instruction Manual Table of ContentsCalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41Cal

Strany 46

44 • Calibration Eagle Instruction ManualCalibrating with the Single Calibration methodThis section describes calibration using the Single Calibration

Strany 47 - Calibration

Eagle Instruction Manual Calibration • 457. Press the POWER/ENTER button to set the span value. SINGLE CALIBRATION END displays, then the Single Calib

Strany 48 - Calibrating the Eagle

46 • Maintenance Eagle Instruction ManualMaintenanceDisplaying the Battery VoltageCheck the battery voltage periodically. Replace or recharge the batt

Strany 49

Eagle Instruction Manual Maintenance • 47NOTE: Setup mode allows you to select between alkaline and Ni-Cd batteries. The two types of batteries have u

Strany 50

48 • Maintenance Eagle Instruction ManualReplacing the oxygen sensorReplace the oxygen sensor when:• The O2 channel cannot be set to 00.0% on an oxyge

Strany 51

Eagle Instruction Manual Maintenance • 49To replace the H2S or CO sensor:1. Take the Eagle to a non-hazardous location, and turn the power off.2. Unsc

Strany 52 - Maintenance

50 • Appendix A: Parts List Eagle Instruction ManualAppendix A: Parts ListTable 6 lists part numbers for the Eagle’s replacement parts and accessories

Strany 53 - Replacing Sensors

Eagle Instruction Manual Appendix A: Parts List • 5180-0131RK Probe, 10-inch hydrophobic (standard probe)80-0132RK Probe, 10-inch hydrophobic (for non

Strany 54 - Replacing the oxygen sensor

52 • Appendix B: Methane Elimination Eagle Instruction ManualAppendix B: Methane EliminationFor applications where methane is an interfering gas, you

Strany 55

Eagle Instruction Manual Appendix B: Methane Elimination • 532. Allow 2 minutes for the combustibles sensor to stabilize.3. Perform the demand zero pr

Strany 56 - Appendix A: Parts List

Table of Contents Eagle Instruction ManualAppendix F: Eagle Tank Tester Model . . . . . . . . . . . . . . . . . . . . . .67Description. . . . . . . .

Strany 57 - Table 6: Parts List (cont.)

54 • Appendix C: Non-Standard Toxic Gas Sensors Eagle Instruction ManualAppendix C: Non-Standard Toxic Gas SensorsAppendix C describes the Eagle’s non

Strany 58

Eagle Instruction Manual Appendix C: Non-Standard Toxic Gas Sensors • 55to the sensor allows the sample to flow through the non-standard toxic gas sens

Strany 59

56 • Appendix C: Non-Standard Toxic Gas Sensors Eagle Instruction ManualAdjusting the sensor controlsCAUTION:Only perform the following steps if you a

Strany 60 - Specifications

Eagle Instruction Manual Appendix C: Non-Standard Toxic Gas Sensors • 57NOTE: If a non-standard toxics channel displays Zero Fail after the Demand Zer

Strany 61

58 • Appendix C: Non-Standard Toxic Gas Sensors Eagle Instruction ManualNOTE: Allow up to 4 hours after you replace a non-standard toxic gas sensor, o

Strany 62 - Adjusting the sensor controls

Eagle Instruction Manual Appendix D: Carbon Dioxide Sensors • 59Appendix D: Carbon Dioxide SensorsAppendix D describes the Eagle’s non-standard, infra

Strany 63

60 • Appendix D: Carbon Dioxide Sensors Eagle Instruction ManualNormal Operation of Carbon Dioxide SensorsCarbon dioxide is a background gas in fresh

Strany 64 - Parts List

Eagle Instruction Manual Appendix D: Carbon Dioxide Sensors • 61NOTE: If you are using the Auto Calibration method, go the Calibration Values screen,

Strany 65

62 • Appendix D: Carbon Dioxide Sensors Eagle Instruction ManualCO2 to a maximum of 3.00% CO2, set the display to 2.00% CO2.2. Unscrew the two large s

Strany 66

Eagle Instruction Manual Appendix D: Carbon Dioxide Sensors • 63Parts ListTable 12 lists part numbers for replacement parts and accessories of the Eag

Strany 67

Eagle Instruction Manual Introduction • 1IntroductionOverviewThe RKI Eagle is the most advanced portable gas detection system available. The Eagle is

Strany 68

64 • Appendix E: Infrared Methane Sensors Eagle Instruction ManualAppendix E: Infrared Methane SensorsAppendix E describes the Eagle’s non-standard, i

Strany 69

Eagle Instruction Manual Appendix E: Infrared Methane Sensors • 65DescriptionThe infrared methane sensor is mounted in the front half of the instrumen

Strany 70 - Target Gases

66 • Appendix E: Infrared Methane Sensors Eagle Instruction Manual1. Use the AIR/▲ and SHIFT/▼ buttons to set the reading to the middle of the range i

Strany 71

Eagle Instruction Manual Appendix F: Eagle Tank Tester Model • 67Appendix F: Eagle Tank Tester ModelThe Eagle Tank Tester model is intended for checki

Strany 72

68 • Appendix F: Eagle Tank Tester Model Eagle Instruction ManualDilution fitting (1:1)CAUTION:When measuring oxygen readings, remove the dilution fitti

Strany 73

Eagle Instruction Manual Appendix F: Eagle Tank Tester Model • 69CalibrationUse a hexane calibrating source to calibrate the combustible gas LEL range

Strany 74 - Dilution fitting (1:1)

70 • Appendix G: Five-Gas and Six-Gas Models Eagle Instruction ManualAppendix G: Five-Gas and Six-Gas ModelsOverviewThe standard Eagle gas monitor inc

Strany 75

Eagle Instruction Manual Appendix G: Five-Gas and Six-Gas Models • 71• Cursors to the left of the first and last channels displayed indicate that addit

Strany 76 - Overview

72 • Appendix H: Eagle Transformer Gas Tester Model Eagle Instruction ManualAppendix H: Eagle Transformer Gas Tester ModelThis Eagle Transformer Gas T

Strany 77

Eagle Instruction Manual Appendix H: Eagle Transformer Gas Tester Model • 73NOTE: The EAGLE can be calibrated either with or without the dilution fitti

Strany 78

2 • Introduction Eagle Instruction ManualAbout this ManualThis manual is intended for use with all Eagle models. Examples used in this manual are for

Strany 79 - Part Number Description

74 • Appendix I: Installing the Data Logger Board Eagle Instruction ManualAppendix I: Installing the Data Logger BoardAppendix H describes the procedu

Strany 80

Eagle Instruction Manual Specifications • 3SpecificationsTable 1 lists physical and environmental specifications for the Eagle. Table 2 lists specificati